1 00:00:20,900 --> 00:00:25,297 This video shows you how to calibrate your OMNIcheck with the Cal-Kit, 2 00:00:25,297 --> 00:00:28,187 from opening the case to sending your results. 3 00:00:28,187 --> 00:00:31,265 Allow about an hour in all, including packing up. 4 00:00:31,265 --> 00:00:34,530 The step numbers match the printed instructions. 5 00:00:36,580 --> 00:00:37,534 First, safety. 6 00:00:37,534 --> 00:00:40,875 This kit uses high-pressure air, up to 5,500 psi. 7 00:00:40,875 --> 00:00:45,920 Wear safety glasses from the moment you connect the air until it's vented. 8 00:00:45,920 --> 00:00:50,419 Keep the air source closed until the software asks you to open it. 9 00:00:50,419 --> 00:00:55,850 And never connect, disconnect or loosen a fitting while it's under pressure. 10 00:00:56,900 --> 00:01:00,992 If you hear a leak, or the pressure keeps rising, press E STOP: 11 00:01:00,992 --> 00:01:04,824 the red button on the kit, or the yellow one on the screen. 12 00:01:04,824 --> 00:01:10,539 E STOP stops the kit and vents the air inside it, but it does not close your air source. 13 00:01:10,539 --> 00:01:16,384 Close the source valve straight away, or your cylinder keeps bleeding out through the kit. 14 00:01:16,384 --> 00:01:21,060 Don't touch any fitting until you've vented and confirmed zero pressure. 15 00:01:21,060 --> 00:01:25,087 To restart, fix the cause, twist the red button to release it, 16 00:01:25,087 --> 00:01:28,000 and start again from Initialize Machines. 17 00:01:29,050 --> 00:01:32,034 Later, the lung inside the headform moves by itself. 18 00:01:32,034 --> 00:01:35,670 Keep your fingers out of the headform mouth while it moves. 19 00:01:39,570 --> 00:01:43,423 You'll need your OMNIcheck with its headform, 20 00:01:43,423 --> 00:01:48,132 its two USB cables, its USB hub and its USB microphone. 21 00:01:48,132 --> 00:01:52,241 A Windows PC with the newest OMNIcheck software. 22 00:01:52,241 --> 00:01:57,720 High-pressure air of at least 4,500 psi, and no more than 5,500. 23 00:01:57,720 --> 00:02:00,203 Two free USB ports on the PC. 24 00:02:00,203 --> 00:02:02,080 And safety glasses. 25 00:02:03,130 --> 00:02:04,801 Here's the whole setup. 26 00:02:04,801 --> 00:02:07,998 Air flows from your source into the Cal-Kit, 27 00:02:07,998 --> 00:02:12,285 which feeds the OMNIcheck at high, medium and low pressure. 28 00:02:12,285 --> 00:02:17,040 The PC runs the software, and talks to both machines over USB. 29 00:02:18,090 --> 00:02:20,449 This is the Cal-Kit's top panel. 30 00:02:20,449 --> 00:02:23,988 The letters A to J are used all the way through. 31 00:02:23,988 --> 00:02:28,780 The air ports are on the left, with the orange low-pressure port. 32 00:02:28,780 --> 00:02:32,790 Power, USB, the switch and E STOP are on the right. 33 00:02:36,620 --> 00:02:38,771 Lay the case flat, wheels down. 34 00:02:38,771 --> 00:02:43,005 It's heavy: one person can lift it, but two people are safer. 35 00:02:43,005 --> 00:02:46,614 Open both latches and lift the lid all the way back. 36 00:02:46,614 --> 00:02:50,820 Keep all the foam: you'll pack the kit back the same way. 37 00:02:51,870 --> 00:02:55,373 Lift out the top foam, with the accessory pouch in it. 38 00:02:55,373 --> 00:03:00,692 Lay the pouch contents on the bench, and check each item against the packing list. 39 00:03:00,692 --> 00:03:04,519 If anything is missing, contact ATOR Labs before you start. 40 00:03:04,519 --> 00:03:09,310 Never use a damaged hose or fitting: it can release high-pressure air. 41 00:03:10,360 --> 00:03:14,746 Below the foam, the two blue HP hoses are coiled around the Cal-Kit. 42 00:03:14,746 --> 00:03:20,930 Lift them out, then lift the Cal-Kit onto the bench beside your OMNIcheck, and open its lid. 43 00:03:25,670 --> 00:03:27,884 Don't start the software yet. 44 00:03:27,884 --> 00:03:33,459 On the back of the OMNIcheck, plug the black cable into the top USB port, 45 00:03:33,459 --> 00:03:36,437 and the grey cable into the bottom one. 46 00:03:36,437 --> 00:03:41,400 Plug both into the OMNIcheck USB hub, and the USB microphone too. 47 00:03:41,400 --> 00:03:43,691 Then plug the hub into the PC. 48 00:03:43,691 --> 00:03:50,028 Last, connect the Cal-Kit's USB-DAQ port, F, to the PC with the supplied USB cable. 49 00:03:50,028 --> 00:03:54,860 In the picture, each cable has the same number at both ends. 50 00:03:55,910 --> 00:04:00,905 Plug the 24 V supply into the jack marked E, then into the wall. 51 00:04:00,905 --> 00:04:07,320 Switch the Cal-Kit on at G, check that E STOP is up, and turn the OMNIcheck on. 52 00:04:08,370 --> 00:04:10,583 Check the five USB connections. 53 00:04:10,583 --> 00:04:12,939 Then open the OMNIcheck software, 54 00:04:12,939 --> 00:04:17,580 and wait until OMNI Connected and Cal Kit Connected are both lit. 55 00:04:17,580 --> 00:04:20,222 Click Auto Calibration, then I Agree. 56 00:04:20,222 --> 00:04:22,150 No special login is needed. 57 00:04:22,150 --> 00:04:24,970 Keep cables and hoses off the floor. 58 00:04:29,350 --> 00:04:31,085 Before you connect anything, 59 00:04:31,085 --> 00:04:35,300 check that the air source valve and the bleed valve are both closed. 60 00:04:35,300 --> 00:04:37,221 Then click Initialize Machines. 61 00:04:37,221 --> 00:04:40,570 The software now asks for each connection in turn. 62 00:04:41,620 --> 00:04:44,367 Take the blue hose tagged HP to OMNI. 63 00:04:44,367 --> 00:04:47,782 Connect its quick-disconnect end to HP out, B: 64 00:04:47,782 --> 00:04:52,384 pull the sleeve back, push the hose on, and let the sleeve go. 65 00:04:52,384 --> 00:04:59,808 Then hand-tighten the nut on the other end onto the OMNIcheck's high-pressure inlet until it's snug. 66 00:04:59,808 --> 00:05:04,113 That's the silver fitting, low on the front, on the right. 67 00:05:04,113 --> 00:05:09,310 Then the clear MP hose: connect its quick-disconnect end to MP out, C, 68 00:05:09,310 --> 00:05:14,803 and the other end to the medium-pressure inlet, on the left, the same way. 69 00:05:14,803 --> 00:05:18,812 Each hose fits only its own inlet, so never force one. 70 00:05:18,812 --> 00:05:22,895 Pull back gently on every connection: it must stay put. 71 00:05:22,895 --> 00:05:26,040 Connect both hoses before you click OK. 72 00:05:27,090 --> 00:05:30,178 Check again that the air source is closed. 73 00:05:30,178 --> 00:05:36,428 Connect the blue supply hose elbow to HP in, port A: sleeve back, push it on, let go. 74 00:05:36,428 --> 00:05:40,399 Fit the supply adapter to your air source, hand-tight, 75 00:05:40,399 --> 00:05:44,516 with its bleed valve closed, and connect the hose to it. 76 00:05:44,516 --> 00:05:50,325 If your source is an upright cylinder, chain or strap it before you connect it. 77 00:05:50,325 --> 00:05:53,590 Never use a source rated above 5,500 psi. 78 00:05:54,640 --> 00:05:59,099 The clear LP hose set is one line, with a Y-piece near the head. 79 00:05:59,099 --> 00:06:04,254 Push its single end firmly into the orange LP out port, D, until it stops. 80 00:06:04,254 --> 00:06:10,315 Fit the mouth fitting into the headform mouth, and the eye fitting onto the eye sensor. 81 00:06:10,315 --> 00:06:12,196 Give each one a gentle tug. 82 00:06:12,196 --> 00:06:13,659 It must not come out. 83 00:06:13,659 --> 00:06:15,790 Don't kink the clear hoses. 84 00:06:16,840 --> 00:06:18,815 Put on your safety glasses. 85 00:06:18,815 --> 00:06:24,010 When the software says Open HP Setup, open the air source valve slowly. 86 00:06:24,010 --> 00:06:27,302 Supply pressure must read at least 4,500 psi. 87 00:06:27,302 --> 00:06:29,424 Type your name, and click OK. 88 00:06:29,424 --> 00:06:32,204 If you hear an air leak, press E STOP, 89 00:06:32,204 --> 00:06:37,210 then close the source valve at once, and don't touch any fitting. 90 00:06:41,040 --> 00:06:44,732 Click LP Leak Checks, and wait for it to finish. 91 00:06:44,732 --> 00:06:46,270 Then MP Leak Checks. 92 00:06:46,270 --> 00:06:49,520 Each button turns green when it passes. 93 00:06:50,570 --> 00:06:52,614 The HP leak check is optional. 94 00:06:52,614 --> 00:06:56,498 Continue fills the kit once to full pressure, then vents. 95 00:06:56,498 --> 00:07:00,587 If you continue, wait for High Pressure Leak Check Complete. 96 00:07:00,587 --> 00:07:03,040 Skip saves air on a single cylinder, 97 00:07:03,040 --> 00:07:07,810 but a leak may go undetected and stop the calibration from succeeding. 98 00:07:07,810 --> 00:07:10,990 While it fills, stay clear of the fittings. 99 00:07:15,100 --> 00:07:16,081 Click Cal Mic. 100 00:07:16,081 --> 00:07:21,894 Push the microphone fully into the sound calibrator, set it to 94 dB, and click OK. 101 00:07:21,894 --> 00:07:24,836 Hold it there until the software finishes. 102 00:07:24,836 --> 00:07:31,390 Check that Cal Mic turns green, then switch the calibrator off and put it back in its bag. 103 00:07:32,440 --> 00:07:33,547 Click Cal Baro. 104 00:07:33,547 --> 00:07:34,876 It runs by itself. 105 00:07:34,876 --> 00:07:39,850 Wait for the complete message, and check that Cal Baro is green. 106 00:07:40,900 --> 00:07:42,392 Click Cal LP Sensors. 107 00:07:42,392 --> 00:07:47,435 The pressure steps up and down twice: once to calibrate, once to check. 108 00:07:47,435 --> 00:07:49,992 Don't touch the hoses while it runs. 109 00:07:49,992 --> 00:07:58,340 When it's complete, take the mouth fitting out of the mouth and the eye fitting off the eye sensor, then click OK. 110 00:07:59,390 --> 00:08:00,945 Click Cal MP Sensor. 111 00:08:00,945 --> 00:08:03,745 The pressure steps up and back down. 112 00:08:03,745 --> 00:08:09,360 Wait for the complete message, and check that Cal MP Sensor is green. 113 00:08:10,410 --> 00:08:11,842 Click Cal HP Sensor. 114 00:08:11,842 --> 00:08:16,208 The kit goes to full pressure, so stay clear of the fittings. 115 00:08:16,208 --> 00:08:18,356 Wait for the complete message. 116 00:08:18,356 --> 00:08:23,152 When all five sensor buttons are green, Sensor Cal Status shows ON. 117 00:08:23,152 --> 00:08:28,270 If one turns red, it hasn't passed: see the troubleshooting section. 118 00:08:32,170 --> 00:08:33,635 Click ISO Leak Check. 119 00:08:33,635 --> 00:08:39,078 Make sure nothing is in the headform mouth, then click OK, and Position Motor. 120 00:08:39,078 --> 00:08:41,939 The lung moves, so keep your fingers out. 121 00:08:41,939 --> 00:08:47,033 When it stops, push the straight mouth plug in until its collar stops it. 122 00:08:47,033 --> 00:08:48,638 It can't go in too far. 123 00:08:48,638 --> 00:08:52,127 Only if it won't seal, use the green tapered plug, 124 00:08:52,127 --> 00:08:57,290 and stop as soon as it seals: pushed too far, it blocks the sensing holes. 125 00:08:57,290 --> 00:08:58,128 Click Start. 126 00:08:58,128 --> 00:09:03,291 The test pulls a vacuum, settles for a minute, then measures for a minute. 127 00:09:03,291 --> 00:09:06,711 When it asks if the system is open to atmosphere, 128 00:09:06,711 --> 00:09:11,316 wait for the lung to stop, take the mouth plug out, then click OK. 129 00:09:11,316 --> 00:09:14,456 When asked about the exhale valve, answer No. 130 00:09:14,456 --> 00:09:16,590 Wait for Leak Check Passed. 131 00:09:17,640 --> 00:09:22,289 With the lung stopped and the mouth plug out, click Orifice Check. 132 00:09:22,289 --> 00:09:27,432 When asked, push orifice A, marked A, into the mouth with a slight twist. 133 00:09:27,432 --> 00:09:32,716 A compliance warning asks you to make sure the headform is open to the air. 134 00:09:32,716 --> 00:09:38,422 Check that only the orifice is in the mouth and nothing blocks it, then click OK. 135 00:09:38,422 --> 00:09:42,719 If you can't confirm that, don't click OK: contact ATOR Labs. 136 00:09:42,719 --> 00:09:49,450 Then, with the lung stopped, change to orifice B, and follow the messages until it finishes. 137 00:09:53,830 --> 00:09:58,821 Check that the sensor buttons, ISO Leak Check and Orifice Check are green, 138 00:09:58,821 --> 00:10:00,980 and Overall Cal Status shows ON. 139 00:10:00,980 --> 00:10:04,400 HP Leak Checks can stay grey if you skipped it. 140 00:10:05,450 --> 00:10:07,344 Click Finalize, then OK. 141 00:10:07,344 --> 00:10:09,710 The Cal Sessions folder opens. 142 00:10:09,710 --> 00:10:14,129 Pick the file for this run: the newest by date modified, 143 00:10:14,129 --> 00:10:18,310 with your OMNIcheck's serial number and today's date. 144 00:10:18,310 --> 00:10:22,176 Email it, unchanged, to calibration@atorlabs.com. 145 00:10:22,176 --> 00:10:25,030 ATOR Labs sends your certificate. 146 00:10:29,140 --> 00:10:31,930 To shut down, close the air source valve. 147 00:10:31,930 --> 00:10:37,578 Then open the bleed valve slowly, keeping your face and hands away from the outlet. 148 00:10:37,578 --> 00:10:41,932 Wait until the air stops venting and supply pressure reads zero, 149 00:10:41,932 --> 00:10:44,042 and leave the bleed valve open. 150 00:10:44,042 --> 00:10:48,260 Now stop and check: watch supply pressure for one full minute. 151 00:10:48,260 --> 00:10:52,139 It must stay at zero, with no air leaking at any fitting. 152 00:10:52,139 --> 00:10:56,630 If the reading rises, or you hear air, leave everything connected, 153 00:10:56,630 --> 00:11:00,010 keep the source closed, and contact ATOR Labs. 154 00:11:01,060 --> 00:11:06,140 Close the software, switch the Cal-Kit off, and turn the OMNIcheck off. 155 00:11:07,190 --> 00:11:11,590 Only after that one-minute check, disconnect in this order. 156 00:11:11,590 --> 00:11:15,841 To release a quick-connect, push the two halves together, 157 00:11:15,841 --> 00:11:19,197 pull the sleeve back, then pull the hose off. 158 00:11:19,197 --> 00:11:25,163 First, the supply hose at the adapter, then take the adapter off the air source. 159 00:11:25,163 --> 00:11:26,580 The elbow at HP in. 160 00:11:26,580 --> 00:11:31,652 Unscrew the HP hose from the OMNIcheck, then release it from HP out. 161 00:11:31,652 --> 00:11:33,516 The MP hose at both ends. 162 00:11:33,516 --> 00:11:38,811 The LP hose set at port D: press the orange ring down, and pull it out. 163 00:11:38,811 --> 00:11:43,882 Then the USB cables, and last, the 24 V supply: from the wall first. 164 00:11:43,882 --> 00:11:48,980 If a connection won't release, don't force it: contact ATOR Labs. 165 00:11:50,030 --> 00:11:54,834 Put every small part back in the pouch, and tick it off the packing list. 166 00:11:54,834 --> 00:11:58,256 Close the Cal-Kit lid and set it in the bottom foam. 167 00:11:58,256 --> 00:12:01,481 Coil both blue HP hoses in the channel around it, 168 00:12:01,481 --> 00:12:04,903 fit the top foam with the pouch, and latch the case. 169 00:12:04,903 --> 00:12:09,430 Put on the prepaid return label from ATOR Labs, and ship it back. 170 00:12:13,440 --> 00:12:18,936 If you hear an air leak, press E STOP, then close the air source at once. 171 00:12:18,936 --> 00:12:25,864 Don't touch any fitting while the kit is pressurized: vent, and confirm zero pressure first. 172 00:12:25,864 --> 00:12:31,586 Only then re-seat each connection, open the air again, and re-run the check. 173 00:12:31,586 --> 00:12:33,995 If an LP leak check fails again, 174 00:12:33,995 --> 00:12:39,115 find the leak with the diagnostic caps: the printed guide shows how. 175 00:12:39,115 --> 00:12:44,410 For any other check, click Save Diagnostics, and contact ATOR Labs. 176 00:12:45,460 --> 00:12:48,951 If a button turns red, that step hasn't passed. 177 00:12:48,951 --> 00:12:52,219 If you suspect a leak, deal with that first. 178 00:12:52,219 --> 00:12:54,892 Otherwise, run the step again, once. 179 00:12:54,892 --> 00:12:59,970 If it's still red, click Save Diagnostics, and contact ATOR Labs. 180 00:13:01,020 --> 00:13:06,131 If you see USB Device Mismatch or a mic read error, close the software. 181 00:13:06,131 --> 00:13:11,891 Check the microphone is plugged into the hub, and selected as the Windows input. 182 00:13:11,891 --> 00:13:15,275 Then reopen the software and run Cal Mic again. 183 00:13:15,275 --> 00:13:20,780 If it fails a second time, click Save Diagnostics, and contact ATOR Labs. 184 00:13:21,930 --> 00:13:22,621 That's it. 185 00:13:22,621 --> 00:13:25,728 If you have any questions, contact ATOR Labs. 186 00:13:25,728 --> 00:13:28,740 Thanks for calibrating with the Cal-Kit.