WEBVTT

00:00:20.900 --> 00:00:25.297
This video shows you how to calibrate your OMNIcheck with the Cal-Kit,

00:00:25.297 --> 00:00:28.187
from opening the case to sending your results.

00:00:28.187 --> 00:00:31.265
Allow about an hour in all, including packing up.

00:00:31.265 --> 00:00:34.530
The step numbers match the printed instructions.

00:00:36.580 --> 00:00:37.534
First, safety.

00:00:37.534 --> 00:00:40.875
This kit uses high-pressure air, up to 5,500 psi.

00:00:40.875 --> 00:00:45.920
Wear safety glasses from the moment you connect the air until it's vented.

00:00:45.920 --> 00:00:50.419
Keep the air source closed until the software asks you to open it.

00:00:50.419 --> 00:00:55.850
And never connect, disconnect or loosen a fitting while it's under pressure.

00:00:56.900 --> 00:01:00.992
If you hear a leak, or the pressure keeps rising, press E STOP:

00:01:00.992 --> 00:01:04.824
the red button on the kit, or the yellow one on the screen.

00:01:04.824 --> 00:01:10.539
E STOP stops the kit and vents the air inside it, but it does not close your air source.

00:01:10.539 --> 00:01:16.384
Close the source valve straight away, or your cylinder keeps bleeding out through the kit.

00:01:16.384 --> 00:01:21.060
Don't touch any fitting until you've vented and confirmed zero pressure.

00:01:21.060 --> 00:01:25.087
To restart, fix the cause, twist the red button to release it,

00:01:25.087 --> 00:01:28.000
and start again from Initialize Machines.

00:01:29.050 --> 00:01:32.034
Later, the lung inside the headform moves by itself.

00:01:32.034 --> 00:01:35.670
Keep your fingers out of the headform mouth while it moves.

00:01:39.570 --> 00:01:43.423
You'll need your OMNIcheck with its headform,

00:01:43.423 --> 00:01:48.132
its two USB cables, its USB hub and its USB microphone.

00:01:48.132 --> 00:01:52.241
A Windows PC with the newest OMNIcheck software.

00:01:52.241 --> 00:01:57.720
High-pressure air of at least 4,500 psi, and no more than 5,500.

00:01:57.720 --> 00:02:00.203
Two free USB ports on the PC.

00:02:00.203 --> 00:02:02.080
And safety glasses.

00:02:03.130 --> 00:02:04.801
Here's the whole setup.

00:02:04.801 --> 00:02:07.998
Air flows from your source into the Cal-Kit,

00:02:07.998 --> 00:02:12.285
which feeds the OMNIcheck at high, medium and low pressure.

00:02:12.285 --> 00:02:17.040
The PC runs the software, and talks to both machines over USB.

00:02:18.090 --> 00:02:20.449
This is the Cal-Kit's top panel.

00:02:20.449 --> 00:02:23.988
The letters A to J are used all the way through.

00:02:23.988 --> 00:02:28.780
The air ports are on the left, with the orange low-pressure port.

00:02:28.780 --> 00:02:32.790
Power, USB, the switch and E STOP are on the right.

00:02:36.620 --> 00:02:38.771
Lay the case flat, wheels down.

00:02:38.771 --> 00:02:43.005
It's heavy: one person can lift it, but two people are safer.

00:02:43.005 --> 00:02:46.614
Open both latches and lift the lid all the way back.

00:02:46.614 --> 00:02:50.820
Keep all the foam: you'll pack the kit back the same way.

00:02:51.870 --> 00:02:55.373
Lift out the top foam, with the accessory pouch in it.

00:02:55.373 --> 00:03:00.692
Lay the pouch contents on the bench, and check each item against the packing list.

00:03:00.692 --> 00:03:04.519
If anything is missing, contact ATOR Labs before you start.

00:03:04.519 --> 00:03:09.310
Never use a damaged hose or fitting: it can release high-pressure air.

00:03:10.360 --> 00:03:14.746
Below the foam, the two blue HP hoses are coiled around the Cal-Kit.

00:03:14.746 --> 00:03:20.930
Lift them out, then lift the Cal-Kit onto the bench beside your OMNIcheck, and open its lid.

00:03:25.670 --> 00:03:27.884
Don't start the software yet.

00:03:27.884 --> 00:03:33.459
On the back of the OMNIcheck, plug the black cable into the top USB port,

00:03:33.459 --> 00:03:36.437
and the grey cable into the bottom one.

00:03:36.437 --> 00:03:41.400
Plug both into the OMNIcheck USB hub, and the USB microphone too.

00:03:41.400 --> 00:03:43.691
Then plug the hub into the PC.

00:03:43.691 --> 00:03:50.028
Last, connect the Cal-Kit's USB-DAQ port, F, to the PC with the supplied USB cable.

00:03:50.028 --> 00:03:54.860
In the picture, each cable has the same number at both ends.

00:03:55.910 --> 00:04:00.905
Plug the 24 V supply into the jack marked E, then into the wall.

00:04:00.905 --> 00:04:07.320
Switch the Cal-Kit on at G, check that E STOP is up, and turn the OMNIcheck on.

00:04:08.370 --> 00:04:10.583
Check the five USB connections.

00:04:10.583 --> 00:04:12.939
Then open the OMNIcheck software,

00:04:12.939 --> 00:04:17.580
and wait until OMNI Connected and Cal Kit Connected are both lit.

00:04:17.580 --> 00:04:20.222
Click Auto Calibration, then I Agree.

00:04:20.222 --> 00:04:22.150
No special login is needed.

00:04:22.150 --> 00:04:24.970
Keep cables and hoses off the floor.

00:04:29.350 --> 00:04:31.085
Before you connect anything,

00:04:31.085 --> 00:04:35.300
check that the air source valve and the bleed valve are both closed.

00:04:35.300 --> 00:04:37.221
Then click Initialize Machines.

00:04:37.221 --> 00:04:40.570
The software now asks for each connection in turn.

00:04:41.620 --> 00:04:44.367
Take the blue hose tagged HP to OMNI.

00:04:44.367 --> 00:04:47.782
Connect its quick-disconnect end to HP out, B:

00:04:47.782 --> 00:04:52.384
pull the sleeve back, push the hose on, and let the sleeve go.

00:04:52.384 --> 00:04:59.808
Then hand-tighten the nut on the other end onto the OMNIcheck's high-pressure inlet until it's snug.

00:04:59.808 --> 00:05:04.113
That's the silver fitting, low on the front, on the right.

00:05:04.113 --> 00:05:09.310
Then the clear MP hose: connect its quick-disconnect end to MP out, C,

00:05:09.310 --> 00:05:14.803
and the other end to the medium-pressure inlet, on the left, the same way.

00:05:14.803 --> 00:05:18.812
Each hose fits only its own inlet, so never force one.

00:05:18.812 --> 00:05:22.895
Pull back gently on every connection: it must stay put.

00:05:22.895 --> 00:05:26.040
Connect both hoses before you click OK.

00:05:27.090 --> 00:05:30.178
Check again that the air source is closed.

00:05:30.178 --> 00:05:36.428
Connect the blue supply hose elbow to HP in, port A: sleeve back, push it on, let go.

00:05:36.428 --> 00:05:40.399
Fit the supply adapter to your air source, hand-tight,

00:05:40.399 --> 00:05:44.516
with its bleed valve closed, and connect the hose to it.

00:05:44.516 --> 00:05:50.325
If your source is an upright cylinder, chain or strap it before you connect it.

00:05:50.325 --> 00:05:53.590
Never use a source rated above 5,500 psi.

00:05:54.640 --> 00:05:59.099
The clear LP hose set is one line, with a Y-piece near the head.

00:05:59.099 --> 00:06:04.254
Push its single end firmly into the orange LP out port, D, until it stops.

00:06:04.254 --> 00:06:10.315
Fit the mouth fitting into the headform mouth, and the eye fitting onto the eye sensor.

00:06:10.315 --> 00:06:12.196
Give each one a gentle tug.

00:06:12.196 --> 00:06:13.659
It must not come out.

00:06:13.659 --> 00:06:15.790
Don't kink the clear hoses.

00:06:16.840 --> 00:06:18.815
Put on your safety glasses.

00:06:18.815 --> 00:06:24.010
When the software says Open HP Setup, open the air source valve slowly.

00:06:24.010 --> 00:06:27.302
Supply pressure must read at least 4,500 psi.

00:06:27.302 --> 00:06:29.424
Type your name, and click OK.

00:06:29.424 --> 00:06:32.204
If you hear an air leak, press E STOP,

00:06:32.204 --> 00:06:37.210
then close the source valve at once, and don't touch any fitting.

00:06:41.040 --> 00:06:44.732
Click LP Leak Checks, and wait for it to finish.

00:06:44.732 --> 00:06:46.270
Then MP Leak Checks.

00:06:46.270 --> 00:06:49.520
Each button turns green when it passes.

00:06:50.570 --> 00:06:52.614
The HP leak check is optional.

00:06:52.614 --> 00:06:56.498
Continue fills the kit once to full pressure, then vents.

00:06:56.498 --> 00:07:00.587
If you continue, wait for High Pressure Leak Check Complete.

00:07:00.587 --> 00:07:03.040
Skip saves air on a single cylinder,

00:07:03.040 --> 00:07:07.810
but a leak may go undetected and stop the calibration from succeeding.

00:07:07.810 --> 00:07:10.990
While it fills, stay clear of the fittings.

00:07:15.100 --> 00:07:16.081
Click Cal Mic.

00:07:16.081 --> 00:07:21.894
Push the microphone fully into the sound calibrator, set it to 94 dB, and click OK.

00:07:21.894 --> 00:07:24.836
Hold it there until the software finishes.

00:07:24.836 --> 00:07:31.390
Check that Cal Mic turns green, then switch the calibrator off and put it back in its bag.

00:07:32.440 --> 00:07:33.547
Click Cal Baro.

00:07:33.547 --> 00:07:34.876
It runs by itself.

00:07:34.876 --> 00:07:39.850
Wait for the complete message, and check that Cal Baro is green.

00:07:40.900 --> 00:07:42.392
Click Cal LP Sensors.

00:07:42.392 --> 00:07:47.435
The pressure steps up and down twice: once to calibrate, once to check.

00:07:47.435 --> 00:07:49.992
Don't touch the hoses while it runs.

00:07:49.992 --> 00:07:58.340
When it's complete, take the mouth fitting out of the mouth and the eye fitting off the eye sensor, then click OK.

00:07:59.390 --> 00:08:00.945
Click Cal MP Sensor.

00:08:00.945 --> 00:08:03.745
The pressure steps up and back down.

00:08:03.745 --> 00:08:09.360
Wait for the complete message, and check that Cal MP Sensor is green.

00:08:10.410 --> 00:08:11.842
Click Cal HP Sensor.

00:08:11.842 --> 00:08:16.208
The kit goes to full pressure, so stay clear of the fittings.

00:08:16.208 --> 00:08:18.356
Wait for the complete message.

00:08:18.356 --> 00:08:23.152
When all five sensor buttons are green, Sensor Cal Status shows ON.

00:08:23.152 --> 00:08:28.270
If one turns red, it hasn't passed: see the troubleshooting section.

00:08:32.170 --> 00:08:33.635
Click ISO Leak Check.

00:08:33.635 --> 00:08:39.078
Make sure nothing is in the headform mouth, then click OK, and Position Motor.

00:08:39.078 --> 00:08:41.939
The lung moves, so keep your fingers out.

00:08:41.939 --> 00:08:47.033
When it stops, push the straight mouth plug in until its collar stops it.

00:08:47.033 --> 00:08:48.638
It can't go in too far.

00:08:48.638 --> 00:08:52.127
Only if it won't seal, use the green tapered plug,

00:08:52.127 --> 00:08:57.290
and stop as soon as it seals: pushed too far, it blocks the sensing holes.

00:08:57.290 --> 00:08:58.128
Click Start.

00:08:58.128 --> 00:09:03.291
The test pulls a vacuum, settles for a minute, then measures for a minute.

00:09:03.291 --> 00:09:06.711
When it asks if the system is open to atmosphere,

00:09:06.711 --> 00:09:11.316
wait for the lung to stop, take the mouth plug out, then click OK.

00:09:11.316 --> 00:09:14.456
When asked about the exhale valve, answer No.

00:09:14.456 --> 00:09:16.590
Wait for Leak Check Passed.

00:09:17.640 --> 00:09:22.289
With the lung stopped and the mouth plug out, click Orifice Check.

00:09:22.289 --> 00:09:27.432
When asked, push orifice A, marked A, into the mouth with a slight twist.

00:09:27.432 --> 00:09:32.716
A compliance warning asks you to make sure the headform is open to the air.

00:09:32.716 --> 00:09:38.422
Check that only the orifice is in the mouth and nothing blocks it, then click OK.

00:09:38.422 --> 00:09:42.719
If you can't confirm that, don't click OK: contact ATOR Labs.

00:09:42.719 --> 00:09:49.450
Then, with the lung stopped, change to orifice B, and follow the messages until it finishes.

00:09:53.830 --> 00:09:58.821
Check that the sensor buttons, ISO Leak Check and Orifice Check are green,

00:09:58.821 --> 00:10:00.980
and Overall Cal Status shows ON.

00:10:00.980 --> 00:10:04.400
HP Leak Checks can stay grey if you skipped it.

00:10:05.450 --> 00:10:07.344
Click Finalize, then OK.

00:10:07.344 --> 00:10:09.710
The Cal Sessions folder opens.

00:10:09.710 --> 00:10:14.129
Pick the file for this run: the newest by date modified,

00:10:14.129 --> 00:10:18.310
with your OMNIcheck's serial number and today's date.

00:10:18.310 --> 00:10:22.176
Email it, unchanged, to calibration@atorlabs.com.

00:10:22.176 --> 00:10:25.030
ATOR Labs sends your certificate.

00:10:29.140 --> 00:10:31.930
To shut down, close the air source valve.

00:10:31.930 --> 00:10:37.578
Then open the bleed valve slowly, keeping your face and hands away from the outlet.

00:10:37.578 --> 00:10:41.932
Wait until the air stops venting and supply pressure reads zero,

00:10:41.932 --> 00:10:44.042
and leave the bleed valve open.

00:10:44.042 --> 00:10:48.260
Now stop and check: watch supply pressure for one full minute.

00:10:48.260 --> 00:10:52.139
It must stay at zero, with no air leaking at any fitting.

00:10:52.139 --> 00:10:56.630
If the reading rises, or you hear air, leave everything connected,

00:10:56.630 --> 00:11:00.010
keep the source closed, and contact ATOR Labs.

00:11:01.060 --> 00:11:06.140
Close the software, switch the Cal-Kit off, and turn the OMNIcheck off.

00:11:07.190 --> 00:11:11.590
Only after that one-minute check, disconnect in this order.

00:11:11.590 --> 00:11:15.841
To release a quick-connect, push the two halves together,

00:11:15.841 --> 00:11:19.197
pull the sleeve back, then pull the hose off.

00:11:19.197 --> 00:11:25.163
First, the supply hose at the adapter, then take the adapter off the air source.

00:11:25.163 --> 00:11:26.580
The elbow at HP in.

00:11:26.580 --> 00:11:31.652
Unscrew the HP hose from the OMNIcheck, then release it from HP out.

00:11:31.652 --> 00:11:33.516
The MP hose at both ends.

00:11:33.516 --> 00:11:38.811
The LP hose set at port D: press the orange ring down, and pull it out.

00:11:38.811 --> 00:11:43.882
Then the USB cables, and last, the 24 V supply: from the wall first.

00:11:43.882 --> 00:11:48.980
If a connection won't release, don't force it: contact ATOR Labs.

00:11:50.030 --> 00:11:54.834
Put every small part back in the pouch, and tick it off the packing list.

00:11:54.834 --> 00:11:58.256
Close the Cal-Kit lid and set it in the bottom foam.

00:11:58.256 --> 00:12:01.481
Coil both blue HP hoses in the channel around it,

00:12:01.481 --> 00:12:04.903
fit the top foam with the pouch, and latch the case.

00:12:04.903 --> 00:12:09.430
Put on the prepaid return label from ATOR Labs, and ship it back.

00:12:13.440 --> 00:12:18.936
If you hear an air leak, press E STOP, then close the air source at once.

00:12:18.936 --> 00:12:25.864
Don't touch any fitting while the kit is pressurized: vent, and confirm zero pressure first.

00:12:25.864 --> 00:12:31.586
Only then re-seat each connection, open the air again, and re-run the check.

00:12:31.586 --> 00:12:33.995
If an LP leak check fails again,

00:12:33.995 --> 00:12:39.115
find the leak with the diagnostic caps: the printed guide shows how.

00:12:39.115 --> 00:12:44.410
For any other check, click Save Diagnostics, and contact ATOR Labs.

00:12:45.460 --> 00:12:48.951
If a button turns red, that step hasn't passed.

00:12:48.951 --> 00:12:52.219
If you suspect a leak, deal with that first.

00:12:52.219 --> 00:12:54.892
Otherwise, run the step again, once.

00:12:54.892 --> 00:12:59.970
If it's still red, click Save Diagnostics, and contact ATOR Labs.

00:13:01.020 --> 00:13:06.131
If you see USB Device Mismatch or a mic read error, close the software.

00:13:06.131 --> 00:13:11.891
Check the microphone is plugged into the hub, and selected as the Windows input.

00:13:11.891 --> 00:13:15.275
Then reopen the software and run Cal Mic again.

00:13:15.275 --> 00:13:20.780
If it fails a second time, click Save Diagnostics, and contact ATOR Labs.

00:13:21.930 --> 00:13:22.621
That's it.

00:13:22.621 --> 00:13:25.728
If you have any questions, contact ATOR Labs.

00:13:25.728 --> 00:13:28.740
Thanks for calibrating with the Cal-Kit.