1 00:00:20,900 --> 00:00:25,297 This video shows you how to calibrate your OMNIcheck with the Cal-Kit, 2 00:00:25,297 --> 00:00:28,187 from opening the case to sending your results. 3 00:00:28,187 --> 00:00:31,265 Allow about an hour in all, including packing up. 4 00:00:31,265 --> 00:00:34,530 The step numbers match the printed instructions. 5 00:00:36,580 --> 00:00:37,534 First, safety. 6 00:00:37,534 --> 00:00:40,875 This kit uses high-pressure air, up to 5,500 psi. 7 00:00:40,875 --> 00:00:45,920 Wear safety glasses from the moment you connect the air until it's vented. 8 00:00:45,920 --> 00:00:50,419 Keep the air source closed until the software asks you to open it. 9 00:00:50,419 --> 00:00:55,850 And never connect, disconnect or loosen a fitting while it's under pressure. 10 00:00:56,900 --> 00:01:00,992 If you hear a leak, or the pressure keeps rising, press E STOP: 11 00:01:00,992 --> 00:01:04,824 the red button on the kit, or the yellow one on the screen. 12 00:01:04,824 --> 00:01:10,539 E STOP stops the kit and vents the air inside it, but it does not close your air source. 13 00:01:10,539 --> 00:01:16,384 Close the source valve straight away, or your cylinder keeps bleeding out through the kit. 14 00:01:16,384 --> 00:01:21,060 Don't touch any fitting until you've vented and confirmed zero pressure. 15 00:01:21,060 --> 00:01:25,087 To restart, fix the cause, twist the red button to release it, 16 00:01:25,087 --> 00:01:28,000 and start again from Initialize Machines. 17 00:01:29,050 --> 00:01:32,034 Later, the lung inside the headform moves by itself. 18 00:01:32,034 --> 00:01:35,670 Keep your fingers out of the headform mouth while it moves.