This video shows you how to calibrate your OMNIcheck with the Cal-Kit,

00:00:25.297 --> 00:00:28.187
from opening the case to sending your results.

00:00:28.187 --> 00:00:31.265
Allow about an hour in all, including packing up.

00:00:31.265 --> 00:00:34.530
The step numbers match the printed instructions.

00:00:36.580 --> 00:00:37.534
First, safety.

00:00:37.534 --> 00:00:40.875
This kit uses high-pressure air, up to 5,500 psi.

00:00:40.875 --> 00:00:45.920
Wear safety glasses from the moment you connect the air until it's vented.

00:00:45.920 --> 00:00:50.419
Keep the air source closed until the software asks you to open it.

00:00:50.419 --> 00:00:55.850
And never connect, disconnect or loosen a fitting while it's under pressure.

00:00:56.900 --> 00:01:00.992
If you hear a leak, or the pressure keeps rising, press E STOP:

00:01:00.992 --> 00:01:04.824
the red button on the kit, or the yellow one on the screen.

00:01:04.824 --> 00:01:10.539
E STOP stops the kit and vents the air inside it, but it does not close your air source.

00:01:10.539 --> 00:01:16.384
Close the source valve straight away, or your cylinder keeps bleeding out through the kit.

00:01:16.384 --> 00:01:21.060
Don't touch any fitting until you've vented and confirmed zero pressure.

00:01:21.060 --> 00:01:25.087
To restart, fix the cause, twist the red button to release it,

00:01:25.087 --> 00:01:28.000
and start again from Initialize Machines.

00:01:29.050 --> 00:01:32.034
Later, the lung inside the headform moves by itself.

00:01:32.034 --> 00:01:35.670
Keep your fingers out of the headform mouth while it moves.