1 00:00:03,350 --> 00:00:07,203 You'll need your OMNIcheck with its headform, 2 00:00:07,203 --> 00:00:11,912 its two USB cables, its USB hub and its USB microphone. 3 00:00:11,912 --> 00:00:16,021 A Windows PC with the newest OMNIcheck software. 4 00:00:16,021 --> 00:00:21,500 High-pressure air of at least 4,500 psi, and no more than 5,500. 5 00:00:21,500 --> 00:00:23,983 Two free USB ports on the PC. 6 00:00:23,983 --> 00:00:25,860 And safety glasses. 7 00:00:26,910 --> 00:00:28,581 Here's the whole setup. 8 00:00:28,581 --> 00:00:31,778 Air flows from your source into the Cal-Kit, 9 00:00:31,778 --> 00:00:36,065 which feeds the OMNIcheck at high, medium and low pressure. 10 00:00:36,065 --> 00:00:40,820 The PC runs the software, and talks to both machines over USB. 11 00:00:41,870 --> 00:00:44,229 This is the Cal-Kit's top panel. 12 00:00:44,229 --> 00:00:47,768 The letters A to J are used all the way through. 13 00:00:47,768 --> 00:00:52,560 The air ports are on the left, with the orange low-pressure port. 14 00:00:52,560 --> 00:00:56,570 Power, USB, the switch and E STOP are on the right.