1 00:00:04,190 --> 00:00:06,404 Don't start the software yet. 2 00:00:06,404 --> 00:00:11,979 On the back of the OMNIcheck, plug the black cable into the top USB port, 3 00:00:11,979 --> 00:00:14,957 and the grey cable into the bottom one. 4 00:00:14,957 --> 00:00:19,920 Plug both into the OMNIcheck USB hub, and the USB microphone too. 5 00:00:19,920 --> 00:00:22,211 Then plug the hub into the PC. 6 00:00:22,211 --> 00:00:28,548 Last, connect the Cal-Kit's USB-DAQ port, F, to the PC with the supplied USB cable. 7 00:00:28,548 --> 00:00:33,380 In the picture, each cable has the same number at both ends. 8 00:00:34,430 --> 00:00:39,425 Plug the 24 V supply into the jack marked E, then into the wall. 9 00:00:39,425 --> 00:00:45,840 Switch the Cal-Kit on at G, check that E STOP is up, and turn the OMNIcheck on. 10 00:00:46,890 --> 00:00:49,103 Check the five USB connections. 11 00:00:49,103 --> 00:00:51,459 Then open the OMNIcheck software, 12 00:00:51,459 --> 00:00:56,100 and wait until OMNI Connected and Cal Kit Connected are both lit. 13 00:00:56,100 --> 00:00:58,742 Click Auto Calibration, then I Agree. 14 00:00:58,742 --> 00:01:00,670 No special login is needed. 15 00:01:00,670 --> 00:01:03,490 Keep cables and hoses off the floor.