00:00:04.190 --> 00:00:06.404
Don't start the software yet.

00:00:06.404 --> 00:00:11.979
On the back of the OMNIcheck, plug the black cable into the top USB port,

00:00:11.979 --> 00:00:14.957
and the grey cable into the bottom one.

00:00:14.957 --> 00:00:19.920
Plug both into the OMNIcheck USB hub, and the USB microphone too.

00:00:19.920 --> 00:00:22.211
Then plug the hub into the PC.

00:00:22.211 --> 00:00:28.548
Last, connect the Cal-Kit's USB-DAQ port, F, to the PC with the supplied USB cable.

00:00:28.548 --> 00:00:33.380
In the picture, each cable has the same number at both ends.

00:00:34.430 --> 00:00:39.425
Plug the 24 V supply into the jack marked E, then into the wall.

00:00:39.425 --> 00:00:45.840
Switch the Cal-Kit on at G, check that E STOP is up, and turn the OMNIcheck on.

00:00:46.890 --> 00:00:49.103
Check the five USB connections.

00:00:49.103 --> 00:00:51.459
Then open the OMNIcheck software,

00:00:51.459 --> 00:00:56.100
and wait until OMNI Connected and Cal Kit Connected are both lit.

00:00:56.100 --> 00:00:58.742
Click Auto Calibration, then I Agree.

00:00:58.742 --> 00:01:00.670
No special login is needed.

00:01:00.670 --> 00:01:03.490
Keep cables and hoses off the floor.