1 00:00:03,830 --> 00:00:05,565 Before you connect anything, 2 00:00:05,565 --> 00:00:09,780 check that the air source valve and the bleed valve are both closed. 3 00:00:09,780 --> 00:00:11,701 Then click Initialize Machines. 4 00:00:11,701 --> 00:00:15,050 The software now asks for each connection in turn. 5 00:00:16,100 --> 00:00:18,847 Take the blue hose tagged HP to OMNI. 6 00:00:18,847 --> 00:00:22,262 Connect its quick-disconnect end to HP out, B: 7 00:00:22,262 --> 00:00:26,864 pull the sleeve back, push the hose on, and let the sleeve go. 8 00:00:26,864 --> 00:00:34,288 Then hand-tighten the nut on the other end onto the OMNIcheck's high-pressure inlet until it's snug. 9 00:00:34,288 --> 00:00:38,593 That's the silver fitting, low on the front, on the right. 10 00:00:38,593 --> 00:00:43,790 Then the clear MP hose: connect its quick-disconnect end to MP out, C, 11 00:00:43,790 --> 00:00:49,283 and the other end to the medium-pressure inlet, on the left, the same way. 12 00:00:49,283 --> 00:00:53,292 Each hose fits only its own inlet, so never force one. 13 00:00:53,292 --> 00:00:57,375 Pull back gently on every connection: it must stay put. 14 00:00:57,375 --> 00:01:00,520 Connect both hoses before you click OK. 15 00:01:01,570 --> 00:01:04,658 Check again that the air source is closed. 16 00:01:04,658 --> 00:01:10,908 Connect the blue supply hose elbow to HP in, port A: sleeve back, push it on, let go. 17 00:01:10,908 --> 00:01:14,879 Fit the supply adapter to your air source, hand-tight, 18 00:01:14,879 --> 00:01:18,996 with its bleed valve closed, and connect the hose to it. 19 00:01:18,996 --> 00:01:24,805 If your source is an upright cylinder, chain or strap it before you connect it. 20 00:01:24,805 --> 00:01:28,070 Never use a source rated above 5,500 psi. 21 00:01:29,120 --> 00:01:33,579 The clear LP hose set is one line, with a Y-piece near the head. 22 00:01:33,579 --> 00:01:38,734 Push its single end firmly into the orange LP out port, D, until it stops. 23 00:01:38,734 --> 00:01:44,795 Fit the mouth fitting into the headform mouth, and the eye fitting onto the eye sensor. 24 00:01:44,795 --> 00:01:46,676 Give each one a gentle tug. 25 00:01:46,676 --> 00:01:48,139 It must not come out. 26 00:01:48,139 --> 00:01:50,270 Don't kink the clear hoses. 27 00:01:51,320 --> 00:01:53,295 Put on your safety glasses. 28 00:01:53,295 --> 00:01:58,490 When the software says Open HP Setup, open the air source valve slowly. 29 00:01:58,490 --> 00:02:01,782 Supply pressure must read at least 4,500 psi. 30 00:02:01,782 --> 00:02:03,904 Type your name, and click OK. 31 00:02:03,904 --> 00:02:06,684 If you hear an air leak, press E STOP, 32 00:02:06,684 --> 00:02:11,690 then close the source valve at once, and don't touch any fitting.