00:00:03.830 --> 00:00:05.565
Before you connect anything,

00:00:05.565 --> 00:00:09.780
check that the air source valve and the bleed valve are both closed.

00:00:09.780 --> 00:00:11.701
Then click Initialize Machines.

00:00:11.701 --> 00:00:15.050
The software now asks for each connection in turn.

00:00:16.100 --> 00:00:18.847
Take the blue hose tagged HP to OMNI.

00:00:18.847 --> 00:00:22.262
Connect its quick-disconnect end to HP out, B:

00:00:22.262 --> 00:00:26.864
pull the sleeve back, push the hose on, and let the sleeve go.

00:00:26.864 --> 00:00:34.288
Then hand-tighten the nut on the other end onto the OMNIcheck's high-pressure inlet until it's snug.

00:00:34.288 --> 00:00:38.593
That's the silver fitting, low on the front, on the right.

00:00:38.593 --> 00:00:43.790
Then the clear MP hose: connect its quick-disconnect end to MP out, C,

00:00:43.790 --> 00:00:49.283
and the other end to the medium-pressure inlet, on the left, the same way.

00:00:49.283 --> 00:00:53.292
Each hose fits only its own inlet, so never force one.

00:00:53.292 --> 00:00:57.375
Pull back gently on every connection: it must stay put.

00:00:57.375 --> 00:01:00.520
Connect both hoses before you click OK.

00:01:01.570 --> 00:01:04.658
Check again that the air source is closed.

00:01:04.658 --> 00:01:10.908
Connect the blue supply hose elbow to HP in, port A: sleeve back, push it on, let go.

00:01:10.908 --> 00:01:14.879
Fit the supply adapter to your air source, hand-tight,

00:01:14.879 --> 00:01:18.996
with its bleed valve closed, and connect the hose to it.

00:01:18.996 --> 00:01:24.805
If your source is an upright cylinder, chain or strap it before you connect it.

00:01:24.805 --> 00:01:28.070
Never use a source rated above 5,500 psi.

00:01:29.120 --> 00:01:33.579
The clear LP hose set is one line, with a Y-piece near the head.

00:01:33.579 --> 00:01:38.734
Push its single end firmly into the orange LP out port, D, until it stops.

00:01:38.734 --> 00:01:44.795
Fit the mouth fitting into the headform mouth, and the eye fitting onto the eye sensor.

00:01:44.795 --> 00:01:46.676
Give each one a gentle tug.

00:01:46.676 --> 00:01:48.139
It must not come out.

00:01:48.139 --> 00:01:50.270
Don't kink the clear hoses.

00:01:51.320 --> 00:01:53.295
Put on your safety glasses.

00:01:53.295 --> 00:01:58.490
When the software says Open HP Setup, open the air source valve slowly.

00:01:58.490 --> 00:02:01.782
Supply pressure must read at least 4,500 psi.

00:02:01.782 --> 00:02:03.904
Type your name, and click OK.

00:02:03.904 --> 00:02:06.684
If you hear an air leak, press E STOP,

00:02:06.684 --> 00:02:11.690
then close the source valve at once, and don't touch any fitting.