1 00:00:03,560 --> 00:00:04,541 Click Cal Mic. 2 00:00:04,541 --> 00:00:10,354 Push the microphone fully into the sound calibrator, set it to 94 dB, and click OK. 3 00:00:10,354 --> 00:00:13,296 Hold it there until the software finishes. 4 00:00:13,296 --> 00:00:19,850 Check that Cal Mic turns green, then switch the calibrator off and put it back in its bag. 5 00:00:20,900 --> 00:00:22,007 Click Cal Baro. 6 00:00:22,007 --> 00:00:23,336 It runs by itself. 7 00:00:23,336 --> 00:00:28,310 Wait for the complete message, and check that Cal Baro is green. 8 00:00:29,360 --> 00:00:30,852 Click Cal LP Sensors. 9 00:00:30,852 --> 00:00:35,895 The pressure steps up and down twice: once to calibrate, once to check. 10 00:00:35,895 --> 00:00:38,452 Don't touch the hoses while it runs. 11 00:00:38,452 --> 00:00:46,800 When it's complete, take the mouth fitting out of the mouth and the eye fitting off the eye sensor, then click OK. 12 00:00:47,850 --> 00:00:49,405 Click Cal MP Sensor. 13 00:00:49,405 --> 00:00:52,205 The pressure steps up and back down. 14 00:00:52,205 --> 00:00:57,820 Wait for the complete message, and check that Cal MP Sensor is green. 15 00:00:58,870 --> 00:01:00,302 Click Cal HP Sensor. 16 00:01:00,302 --> 00:01:04,668 The kit goes to full pressure, so stay clear of the fittings. 17 00:01:04,668 --> 00:01:06,816 Wait for the complete message. 18 00:01:06,816 --> 00:01:11,612 When all five sensor buttons are green, Sensor Cal Status shows ON. 19 00:01:11,612 --> 00:01:16,730 If one turns red, it hasn't passed: see the troubleshooting section.