Click Cal Mic.

00:00:04.541 --> 00:00:10.354
Push the microphone fully into the sound calibrator, set it to 94 dB, and click OK.

00:00:10.354 --> 00:00:13.296
Hold it there until the software finishes.

00:00:13.296 --> 00:00:19.850
Check that Cal Mic turns green, then switch the calibrator off and put it back in its bag.

00:00:20.900 --> 00:00:22.007
Click Cal Baro.

00:00:22.007 --> 00:00:23.336
It runs by itself.

00:00:23.336 --> 00:00:28.310
Wait for the complete message, and check that Cal Baro is green.

00:00:29.360 --> 00:00:30.852
Click Cal LP Sensors.

00:00:30.852 --> 00:00:35.895
The pressure steps up and down twice: once to calibrate, once to check.

00:00:35.895 --> 00:00:38.452
Don't touch the hoses while it runs.

00:00:38.452 --> 00:00:46.800
When it's complete, take the mouth fitting out of the mouth and the eye fitting off the eye sensor, then click OK.

00:00:47.850 --> 00:00:49.405
Click Cal MP Sensor.

00:00:49.405 --> 00:00:52.205
The pressure steps up and back down.

00:00:52.205 --> 00:00:57.820
Wait for the complete message, and check that Cal MP Sensor is green.

00:00:58.870 --> 00:01:00.302
Click Cal HP Sensor.

00:01:00.302 --> 00:01:04.668
The kit goes to full pressure, so stay clear of the fittings.

00:01:04.668 --> 00:01:06.816
Wait for the complete message.

00:01:06.816 --> 00:01:11.612
When all five sensor buttons are green, Sensor Cal Status shows ON.

00:01:11.612 --> 00:01:16.730
If one turns red, it hasn't passed: see the troubleshooting section.