1 00:00:03,350 --> 00:00:04,815 Click ISO Leak Check. 2 00:00:04,815 --> 00:00:10,258 Make sure nothing is in the headform mouth, then click OK, and Position Motor. 3 00:00:10,258 --> 00:00:13,119 The lung moves, so keep your fingers out. 4 00:00:13,119 --> 00:00:18,213 When it stops, push the straight mouth plug in until its collar stops it. 5 00:00:18,213 --> 00:00:19,818 It can't go in too far. 6 00:00:19,818 --> 00:00:23,307 Only if it won't seal, use the green tapered plug, 7 00:00:23,307 --> 00:00:28,470 and stop as soon as it seals: pushed too far, it blocks the sensing holes. 8 00:00:28,470 --> 00:00:29,308 Click Start. 9 00:00:29,308 --> 00:00:34,471 The test pulls a vacuum, settles for a minute, then measures for a minute. 10 00:00:34,471 --> 00:00:37,891 When it asks if the system is open to atmosphere, 11 00:00:37,891 --> 00:00:42,496 wait for the lung to stop, take the mouth plug out, then click OK. 12 00:00:42,496 --> 00:00:45,636 When asked about the exhale valve, answer No. 13 00:00:45,636 --> 00:00:47,770 Wait for Leak Check Passed. 14 00:00:48,820 --> 00:00:53,469 With the lung stopped and the mouth plug out, click Orifice Check. 15 00:00:53,469 --> 00:00:58,612 When asked, push orifice A, marked A, into the mouth with a slight twist. 16 00:00:58,612 --> 00:01:03,896 A compliance warning asks you to make sure the headform is open to the air. 17 00:01:03,896 --> 00:01:09,602 Check that only the orifice is in the mouth and nothing blocks it, then click OK. 18 00:01:09,602 --> 00:01:13,899 If you can't confirm that, don't click OK: contact ATOR Labs. 19 00:01:13,899 --> 00:01:20,630 Then, with the lung stopped, change to orifice B, and follow the messages until it finishes.