WEBVTT

00:00:03.350 --> 00:00:04.815
Click ISO Leak Check.

00:00:04.815 --> 00:00:10.258
Make sure nothing is in the headform mouth, then click OK, and Position Motor.

00:00:10.258 --> 00:00:13.119
The lung moves, so keep your fingers out.

00:00:13.119 --> 00:00:18.213
When it stops, push the straight mouth plug in until its collar stops it.

00:00:18.213 --> 00:00:19.818
It can't go in too far.

00:00:19.818 --> 00:00:23.307
Only if it won't seal, use the green tapered plug,

00:00:23.307 --> 00:00:28.470
and stop as soon as it seals: pushed too far, it blocks the sensing holes.

00:00:28.470 --> 00:00:29.308
Click Start.

00:00:29.308 --> 00:00:34.471
The test pulls a vacuum, settles for a minute, then measures for a minute.

00:00:34.471 --> 00:00:37.891
When it asks if the system is open to atmosphere,

00:00:37.891 --> 00:00:42.496
wait for the lung to stop, take the mouth plug out, then click OK.

00:00:42.496 --> 00:00:45.636
When asked about the exhale valve, answer No.

00:00:45.636 --> 00:00:47.770
Wait for Leak Check Passed.

00:00:48.820 --> 00:00:53.469
With the lung stopped and the mouth plug out, click Orifice Check.

00:00:53.469 --> 00:00:58.612
When asked, push orifice A, marked A, into the mouth with a slight twist.

00:00:58.612 --> 00:01:03.896
A compliance warning asks you to make sure the headform is open to the air.

00:01:03.896 --> 00:01:09.602
Check that only the orifice is in the mouth and nothing blocks it, then click OK.

00:01:09.602 --> 00:01:13.899
If you can't confirm that, don't click OK: contact ATOR Labs.

00:01:13.899 --> 00:01:20.630
Then, with the lung stopped, change to orifice B, and follow the messages until it finishes.