1 00:00:03,560 --> 00:00:06,350 To shut down, close the air source valve. 2 00:00:06,350 --> 00:00:11,998 Then open the bleed valve slowly, keeping your face and hands away from the outlet. 3 00:00:11,998 --> 00:00:16,352 Wait until the air stops venting and supply pressure reads zero, 4 00:00:16,352 --> 00:00:18,462 and leave the bleed valve open. 5 00:00:18,462 --> 00:00:22,680 Now stop and check: watch supply pressure for one full minute. 6 00:00:22,680 --> 00:00:26,559 It must stay at zero, with no air leaking at any fitting. 7 00:00:26,559 --> 00:00:31,050 If the reading rises, or you hear air, leave everything connected, 8 00:00:31,050 --> 00:00:34,430 keep the source closed, and contact ATOR Labs. 9 00:00:35,480 --> 00:00:40,560 Close the software, switch the Cal-Kit off, and turn the OMNIcheck off. 10 00:00:41,610 --> 00:00:46,010 Only after that one-minute check, disconnect in this order. 11 00:00:46,010 --> 00:00:50,261 To release a quick-connect, push the two halves together, 12 00:00:50,261 --> 00:00:53,617 pull the sleeve back, then pull the hose off. 13 00:00:53,617 --> 00:00:59,583 First, the supply hose at the adapter, then take the adapter off the air source. 14 00:00:59,583 --> 00:01:01,000 The elbow at HP in. 15 00:01:01,000 --> 00:01:06,072 Unscrew the HP hose from the OMNIcheck, then release it from HP out. 16 00:01:06,072 --> 00:01:07,936 The MP hose at both ends. 17 00:01:07,936 --> 00:01:13,231 The LP hose set at port D: press the orange ring down, and pull it out. 18 00:01:13,231 --> 00:01:18,302 Then the USB cables, and last, the 24 V supply: from the wall first. 19 00:01:18,302 --> 00:01:23,400 If a connection won't release, don't force it: contact ATOR Labs. 20 00:01:24,450 --> 00:01:29,254 Put every small part back in the pouch, and tick it off the packing list. 21 00:01:29,254 --> 00:01:32,676 Close the Cal-Kit lid and set it in the bottom foam. 22 00:01:32,676 --> 00:01:35,901 Coil both blue HP hoses in the channel around it, 23 00:01:35,901 --> 00:01:39,323 fit the top foam with the pouch, and latch the case. 24 00:01:39,323 --> 00:01:43,850 Put on the prepaid return label from ATOR Labs, and ship it back.