1 00:00:03,460 --> 00:00:08,956 If you hear an air leak, press E STOP, then close the air source at once. 2 00:00:08,956 --> 00:00:15,884 Don't touch any fitting while the kit is pressurized: vent, and confirm zero pressure first. 3 00:00:15,884 --> 00:00:21,606 Only then re-seat each connection, open the air again, and re-run the check. 4 00:00:21,606 --> 00:00:24,015 If an LP leak check fails again, 5 00:00:24,015 --> 00:00:29,135 find the leak with the diagnostic caps: the printed guide shows how. 6 00:00:29,135 --> 00:00:34,430 For any other check, click Save Diagnostics, and contact ATOR Labs. 7 00:00:35,480 --> 00:00:38,971 If a button turns red, that step hasn't passed. 8 00:00:38,971 --> 00:00:42,239 If you suspect a leak, deal with that first. 9 00:00:42,239 --> 00:00:44,912 Otherwise, run the step again, once. 10 00:00:44,912 --> 00:00:49,990 If it's still red, click Save Diagnostics, and contact ATOR Labs. 11 00:00:51,040 --> 00:00:56,151 If you see USB Device Mismatch or a mic read error, close the software. 12 00:00:56,151 --> 00:01:01,911 Check the microphone is plugged into the hub, and selected as the Windows input. 13 00:01:01,911 --> 00:01:05,295 Then reopen the software and run Cal Mic again. 14 00:01:05,295 --> 00:01:10,800 If it fails a second time, click Save Diagnostics, and contact ATOR Labs. 15 00:01:11,950 --> 00:01:12,641 That's it. 16 00:01:12,641 --> 00:01:15,748 If you have any questions, contact ATOR Labs. 17 00:01:15,748 --> 00:01:18,760 Thanks for calibrating with the Cal-Kit.