If you hear an air leak, press E STOP, then close the air source at once.

00:00:08.956 --> 00:00:15.884
Don't touch any fitting while the kit is pressurized: vent, and confirm zero pressure first.

00:00:15.884 --> 00:00:21.606
Only then re-seat each connection, open the air again, and re-run the check.

00:00:21.606 --> 00:00:24.015
If an LP leak check fails again,

00:00:24.015 --> 00:00:29.135
find the leak with the diagnostic caps: the printed guide shows how.

00:00:29.135 --> 00:00:34.430
For any other check, click Save Diagnostics, and contact ATOR Labs.

00:00:35.480 --> 00:00:38.971
If a button turns red, that step hasn't passed.

00:00:38.971 --> 00:00:42.239
If you suspect a leak, deal with that first.

00:00:42.239 --> 00:00:44.912
Otherwise, run the step again, once.

00:00:44.912 --> 00:00:49.990
If it's still red, click Save Diagnostics, and contact ATOR Labs.

00:00:51.040 --> 00:00:56.151
If you see USB Device Mismatch or a mic read error, close the software.

00:00:56.151 --> 00:01:01.911
Check the microphone is plugged into the hub, and selected as the Windows input.

00:01:01.911 --> 00:01:05.295
Then reopen the software and run Cal Mic again.

00:01:05.295 --> 00:01:10.800
If it fails a second time, click Save Diagnostics, and contact ATOR Labs.

00:01:11.950 --> 00:01:12.641
That's it.

00:01:12.641 --> 00:01:15.748
If you have any questions, contact ATOR Labs.

00:01:15.748 --> 00:01:18.760
Thanks for calibrating with the Cal-Kit.